Overview
M4735A Heartstream XL Defibrillator/Monitor 1-3
1
The M4735A Heartstream XL Defibrillator/Monitor automatically stores crit-
ical events in its internal memory, such as shocks and alarm violations. Addi-
tional events of interest may also be marked for storage. These events can be
printed as they occur or as part of an Event Summary. The M4735A Heart-
stream XL Defibrillator/Monitor also enables you to store data and events on
an optional Data Card for downloading to CodeRunner Web Data Manage-
ment Systems.
The versatile M4735A Heartstream XL Defibrillator/Monitor is highly con-
figurable to better meet the needs of diverse users. The messages and softkeys
vary, depending on how the M4735A Heartstream XL Defibrillator/Monitor
is configured. Be sure to familiarize yourself with your configuration before
using the M4735A Heartstream XL Defibrillator/Monitor (see “Configuring
the Heartstream XL” on page 10-7).
The M4735A Heartstream XL Defibrillator/Monitor is powered by AC power
and a rechargable sealed lead acid (SLA) battery which allows the defibrilla-
tor to charge to 200 joules in less than three seconds. Proper care of your bat-
teries will ensure that they have the energy required to operate the M4735A
Heartstream XL Defibrillator/Monitor and to deliver the appropriate therapy.
(See “Battery Maintenance” on page 11-7.) Similarly, performing the speci-
fied operational checks will ensure that the M4735A Heartstream XL
Defibrillator/Monitor is functioning properly and ready for use. (See “Opera-
tional Checks” on page 11-2.)