EasyManua.ls Logo

Agilent Technologies M4735A - Page 52

Agilent Technologies M4735A
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Defibrillating
3-12 Defibrillating in AED Mode
3. If shock advised, press .
Once charging is complete, the charge tone becomes continuous. Make
sure no one is touching the patient or anything connected to the patient.
Call out clearly and loudly "Stand Clear." Then press to deliver a
shock to the patient.
)LJXUH3UHVV6+2&.'LVSOD\
:$51,1* 'HILEULOODWLRQFXUUHQWFD QFDXVHRSHUDW RURU E\ VWDQGHU LQMXU\'RQRWWRXFKWKH
SDWLHQWRUHTXLSPHQWFRQQHFWHGWRWKHSDWLHQWGXULQJGHILEULOODWLRQ
The defibrillator automatically disarms within 30 seconds if you do not
press .
6+2&.
6+2&.
LEAD
SELECT
SpO2
ON/OFF
HR
ALARM
120
Pads
SHOCK
Stand Clear!
Press SHOCK
150J
Shocks: 0
00:00:14
6+2&.

Table of Contents