ERC Protocol
M4735A Heartstream XL Defibrillator/Monitor 3-19
3
If CPR is needed, press . While paused, a timer indicates the elapsed
time and the total duration of the Pause state, as shown:
)LJXUH'LVSOD\LQ3DXVHG6WDWH
The total pause duration depends on the event preceding the Pause state. If
you entered the Pause state:
l
at the completion of a shock series or shortly after a shock is deliv-
ered, the duration is equal to the 3RVW6KRFN&35 7LPHU configuration set-
ting (the default setting is 60 seconds).
l
when no shock was advised, the duration is equal to the 16$7LPHU
configuration setting, where NSA is an acronym for No Shock
Advised (the default setting is 180 seconds).
3$86(
Shocks: 3
00:01:40
Paused
RESUME
ANALYZE
Timer: 35 / 60
Elapsed Time
Total Pause Duration
in Seconds