Troubleshooting
M4735A Heartstream XL Defibrillator/Monitor 3-21
3
No Shock
Delivered
l
Pads are not properly connected
to the patient.
l
Check pads connection.
Key
Inactive
l
The key pressed only functions
in Manual Mode.
l
The key pressed does not func-
tion during analysis or charging.
l
The key pressed does not func-
tion while in a pause state.
l
Turn the Energy Select knob to
Manual on prior to pressing the
key.
l
Wait for analysis or charging to
complete prior to pressing the
key.
l
Press prior to pressing
the key.
7DEOH7URXEOHVKRRWLQJLQ$('0RGH