Solving Problems
If You Encounter a Problem
Chapter 486
4. Press
Continue to continue with the sequence (or Abort to abort with no lost files).
At the conclusion of the sequence, perform the following steps:
1. Cycle power once again.
Cycling power restores all previously installed options. You should expect to see several
error messages resulting from calibration files being restored from EEPROM.
2. Perform the DCFM/DC
Φ
M calibration.
Refer to the
DCFM/DCΦM Cal softkey description in Chapter 5, “Key Reference,” in this
manual.
3. Agilent Technologies is interested in the circumstances that caused you to have to initiate
this procedure. Please contact us at the appropriate telephone number listed in Table 4-1.
We would like to help you eliminate any repeat occurrences.