Key Reference
M
Chapter 5 183
Meter Address
Supported All
This softkey enables you to select the external power meter GPIB address for use during
the user flatness calibration.
Preset N/A
Range 0
30
Location Page 232
Remarks The meter address is not affected by a preset or by a power cycle.
The power meter is controlled only through a GPIB cable.
Meter Channel A B
Supported All
This softkey toggles between channels A and B when you are using an external power
meter that has two channels.
Preset N/A
Choices A or B
Location Page 232
Remarks The meter channel selection is not affected by a preset or by a power
cycle.
The power meter is controlled only through a GPIB cable.
Meter Timeout
Supported All
This softkey enables you to select the remote timeout length of the external power meter
during user flatness calibration.
Preset N/A
Range 1.0 msec100.0 sec
Location Page 232
Remarks The meter timeout setting is not affected by a preset or by a power
cycle. The power meter is controlled only through a GPIB cable.