Key Reference
S
Chapter 5 217
Sweep Repeat Single Cont
Supported All
This softkey toggles the sweep repetition between single sweep and continuous sweep
types. The sweep and point triggers will then trigger the sweep. See “Point Trigger” on
page 191 and “Sweep Trigger” on page 217.
Preset Cont
Choices Single or Cont
Location Page 245
Remarks None
Sweep Trigger
Supported All
Use this softkey to access a menu of choices that enables you to select a triggering mode
for a sweep. You can choose triggering that, once a sweep is armed, occurs immediately,
is supplied by the GPIB, occurs on either the positive or negative edge of a signal
supplied to the TRIGGER IN connector, or is initiated by the
Trigger hardkey.
Preset Free Run
Choices Free Run Trigger key Bus Ext
Location Page 245
Remarks You can also configure the polarity of the TTL signal output at the
TRIGGER IN connector using the
Trigger In Polarity Neg Pos softkey
on this menu. Refer to “Trigger In Polarity Neg Pos” on page 222.