Key Reference
D
Chapter 5134
Do Cal
Supported All
This softkey enables you to start a flatness calibration using the information from the
user flatness list sweep table.
Preset N/A
Range N/A
Location Page 232
Remarks None
Done
Supported All
Use this softkey to return to the Flatness menu once the calibration routine is complete.
Preset N/A
Range
Ν/Α
Location Page 233
Remarks None
Do Power Search
Supported All
Use this softkey to execute the power search calibration routine one time. Power search
is an internal calibration routine used to achieve calibrated output power when the ALC
is off. You can execute a power search at any time by pressing this softkey.
Preset N/A
Range N/A
Location Page 232
Remarks The routine will execute only if RF is on and ALC is off.