Fundamental Operation
Configuring Analog Modulation (PSG-A Series Only)
Chapter 238
Activating FM
Follow these remaining steps to output the frequency-modulated signal.
1. Press
FM Off On.
The FM annunciator is activated indicating that you have enabled frequency modulation.
2. Press
RF On/Off.
The RF ON annuciator is activated, indicating that the signal is now available at the
RF OUTPUT connector.
Configuring ΦM
Using this procedure, you will learn how to create a phase-modulated RF carrier with the
following characteristics:
RF output frequency set to 20.0 GHz
RF output amplitude set to 0 dBm
Φ
M deviation set to 0.25
π
radians
Φ
M rate set to 30 kHz
Setting the RF Output Frequency
1. Press
Preset.
2. Press
Frequency > 20 > GHz.
The FREQUENCY area of the display now reads 20.000 000 000 00 GHz.
Setting the RF Output Amplitude
Press
Amplitude > 0 > dBm.
The AMPLITUDE area of the display now reads 0.00 dBm.
Setting the
Φ
M Deviation and Rate
1. Press the
FM/ΦM hardkey.
2. Press the
FM ΦM softkey.
The first level menu of ΦM softkeys is displayed.

Table of Contents

Other manuals for Agilent Technologies PSG Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PSG Series and is the answer not in the manual?

Agilent Technologies PSG Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelPSG Series
CategoryPortable Generator
LanguageEnglish

Related product manuals