Fundamental Operation
Configuring Analog Modulation (PSG-A Series Only)
Chapter 2 39
3. Press
ΦMDev > .25 > pi rad.
This changes the
Φ
M deviation to 0.25
π
radians.
4. Press
ΦMRate > 10 > kHz.
This sets the
Φ
M rate to 10 kHz.
The signal generator is now configured to output a 0 dBm, phase-modulated carrier at 20 GHz
with a 0.25
π
radian deviation and 10 kHz rate. The shape of the waveform is a sinewave.
(Notice that sine is the default for the
ΦM Waveform softkey. Press More (1 of 2) to see the
softkey.)
Activating
Φ
M
Follow these remaining steps to output the phase-modulated signal.
1. Press
ΦMOffOn.
The
Φ
M annunciator is activated, indicating that you have enabled phase modulation.
2. Press
RF On/Off.
The RF ON annunciator is activated, indicating that the signal is now available at the
RF OUTPUT connector.
The signal generator is now configured to output a 0 dBm, pulse-modulated carrier at 18 GHz
with a 100-microsecond pulse period and 24-microsecond pulse width. The pulse source is set
to internal free-run. (Notice that internal free-run is the default for the
Pulse Source softkey.)

Table of Contents

Other manuals for Agilent Technologies PSG Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PSG Series and is the answer not in the manual?

Agilent Technologies PSG Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelPSG Series
CategoryPortable Generator
LanguageEnglish

Related product manuals