Key Reference
P
Chapter 5 195
Pulse Delay
Supported PSG-A Series
This softkey enables you to change the pulse delay of the internally generated pulse
modulation source.
Preset 0.0 sec
Range The range is dependent upon the value set for the pulse period.
Internal Free Run:
(pulse period
20 nS) to (pulse period
20 nS)
Internal Triggered & Doublet: 70nS to (pulse period
20 nS)
Location Page 242
Remarks None
Pulse Off On
Supported PSG-A Series
This softkey toggles the operating state for the currently selected pulse source.
When pulse modulation is on, the PULSE annunciator is shown in the display
Preset Off
Choices Off or On
Location Page 242
Remarks The RF carrier is modulated by the enabled modulation only when you
have also set
Mod On/Off to On.