Key Reference
A
Chapter 5 117
Ampl Start
Supported All
This softkey enables you to set the amplitude of the first point in the sweep. When you
press
Ampl Start, amplitude becomes the active function for the first point in the step
sweep.
Preset
135.00 dBm (for instruments with Option 1E1)
Range You can set the value anywhere in the range of the signal generator’s
specified output power.
Location Page 245
Remarks None
Ampl Stop
Supported All
This softkey enables you to set the amplitude of the last point in the sweep. When you
press
Ampl Stop, amplitude becomes the active function for the last point in the step
sweep.
Preset
135.00 dBm (for instruments with Option 1E1)
Range You can set the value anywhere in the range of the signal generators
specified output power.
Location Page 245
Remarks None