Fundamental Operation
Configuring for Remote Control
Chapter 252
Configuring for an RS-232 Interface
1. Press Utility > GPIB/RS-232 LAN > RS-232 Setup.
2. Press
RS-232 Baud Rate.
Press the desired baud rate softkey to set the baud rate.
3. Press
RS-232 Echo Off On.
This toggles the state of the SCPI echoing on the RS-232 connection. Set as desired.
4. Press
Trans/Recv Pace None Xon.
This toggles from no handshaking (
Trans/Recv Pace None) to XON/XOFF handshaking
(
Trans/Recv Pace Xon) when transmitting or receiving data via RS-232. Set as desired.
5. Press
Reset RS-232.
This deletes the data from the RS-232 buffer. Pressing this key will discard any
unprocessed SCPI input received over RS-232.
6. Press
RS-232 Timeout.
This allows you to enter a value in seconds before the instrument generates an RS-232
timeout after not receiving data on the serial bus.
These RS-232 parameters are not affected by an instrument preset or by a power cycle.

Table of Contents

Other manuals for Agilent Technologies PSG Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PSG Series and is the answer not in the manual?

Agilent Technologies PSG Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelPSG Series
CategoryPortable Generator
LanguageEnglish

Related product manuals