Key Reference
A
Chapter 5108
ALC Off On
Supported All
This softkey toggles the operating state of the automatic leveling control (ALC) circuit.
The RF output power of the signal generator is constantly monitored and controlled by
the ALC circuit. Its purpose is to hold output power at the desired level in spite of drift
due to temperature and time.
Preset On
Range Off or On
Location Page 232
Remarks An alternative to setting the ALC off is to set the ALC to a narrow
bandwidth. ALC off is used for pulse widths less than 1.0 usec.
All
Supported All
This softkey enables you to specify that all file types in the user file system be listed.
Preset N/A
Range N/A
Location Page 249
Remarks None
AM
Supported PSG-A Series
Use this hardkey to access a menu of choices that enables you to configure the amplitude
modulation type, path, depth and depth coupling, rate, source, mode, and waveform.
Preset N/A
Range N/A
Location Page 230
Remarks None

Table of Contents

Other manuals for Agilent Technologies PSG Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PSG Series and is the answer not in the manual?

Agilent Technologies PSG Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelPSG Series
CategoryPortable Generator
LanguageEnglish

Related product manuals