Key Reference
F
Chapter 5152
Freq
Supported All
This softkey enables you to set the signal generator to sweep frequency data only. The
amplitude is set at a constant level determined by the
Amplitude hardkey. You can define
a sweep containing both amplitude and frequency information and still choose to sweep
frequency only.
Preset N/A
Range N/A
Location Page 245
Remarks None
Freq Multiplier
Supported All
This softkey enables you to multiply the frequency shown on the display without
changing the frequency output at the RF OUTPUT connector (simulating the frequency
at the output of a harmonic multiplier).
For any multiplier other than 1, the MULT indicator is shown in the frequency area of the
display.
Preset 1
Range
−
100 to +1000 (excluding 0)
Location Page 239
Remarks The displayed frequency value is equal to the actual frequency, minus
the reference frequency, times the multiplier, plus the offset frequency.