Key Reference
E
Chapter 5140
Ext Detector
Supported All
This softkey enables you to select an external diode detector leveling mode. This enables
you to connect a detector and a coupler or splitter to the RF output and level the RF
signal at a point external to the RF output connector.
Preset N/A
Range N/A
Location Page 232
Remarks None
Ext Detector Coupling Factor
Supported All
This softkey enables you to set the external detector coupling factor.
Preset 16.0 dB
Range
−
200.0 dB to 200.0 dB
Location Page 232
Remarks N/A
Ext Impedance 50 Ohm 600 Ohm
Supported All
This softkey toggles the external input impedance between 50 ohms and 600 ohms.
Preset 50 Ohm
Choices 50 Ohm or 600 Ohm
Location Page 230, Page 235, Page 237
Remarks None