  Constant Current mode will maintain the load current at 
the maximum level set by the current limit value, even 
if the maximum power level is exceeded.  This is done 
by reducing the voltage as needed.  As such, the 
voltage will be reduced from the set level down to zero 
depending on the load requirement.  This mode is 
useful for starting up motor or capacitor loads that may 
require a high inrush current.  This mode also has a 
user programmable trip delay. 
  Voltage sensing for regulation and measurement can be 
selected for internal or external.  External voltage 
sensing can compensate for voltage drops caused by 
load cable impedance.  To achieve the best output 
regulation, select external sense and connect the 
voltage sense wires at the load. 
INITIAL SETUP  The initial setup menu can be used to determine the AC 
source settings at power up. CAUTION: The initial setup 
can be used to power up the AC source with the output 
on and a high voltage present at the output. For normal 
situations, this is not recommended due to the potential 
danger to operators. It is recommended that the initial 
voltage be set low and/or the output relay be 
programmed to OFF for most situations. 
LIMIT SETUP  The Limit menu shows the frequency, voltage, and 
current limit capabilities of the AC source. Any attempt 
to program the output beyond these limits will result in 
a “-222 Data Out of Range error”. Note that these limits 
are hardware determined and cannot be changed by 
the user. They are shown for reference only. 
UTILITY 2 
CONFIGURATION  The Configuration menu shows the installed options. 
This screen is for reference only and typically, no fields 
can be changed by the user.