User Manual Rev BE AMETEK Programmable Power
MX Series 219
400 VDC Range + Full-scale: [Applies to MX Series I only] Program + 320.0 volts.
Once the output settings are made, turn on the ALC
mode and allow the output on all phases to settle. Next,
proceed to the output calibration screen. Select the
VOLT +FS parameter and enter the set voltage in the
full-scale calibration window followed by the ENTER
key. A value between 0.9 and 1.1 will appear in the
window. Save this value by pressing the SET key.