4.3 CDMA Transmitter and Receiver Test --- TX and RX Tester Mode
Function keys:
1st page
[Window] F7: Displays a waveform-window setup menu.
[Slot] F7: Displays a waveform corresponding to one slot.
[Leading] F9: Displays a waveform at the leading edge of the burst signal.
[Trailing] F10: Displays a waveform at the trailing edge of the burst signal.
[return] F12: Returns to the previous menu.
[Marker] F8: Displays the Marker menu.
If this key is pressed when Off is selected, Normal is selected.
[Normal] F7: Sets normal marker mode. Enters the marker-position input wait status.
Range: Lower limit to upper limit of horizontal display scale
(unit: symbol)
Resolution: 0.1 symbol
Initial value: Center of the screen
[Off] F8: Sets marker mode to off and clear the marker (initial value).
[return] F12: Returns to the previous menu.
[Storage Mode] F9: Displays the storage mode setup menu.
Either of the following items can be selected:
1st page
[Normal] F7: Displays the normal waveform storage mode setup menu.
[Average] F8: Sets averaging mode.
Measure mode is set to Single.
[Average Count] F9: Sets the averaging count.
2 to 9,999, Resolution: 1, Initial value: 10
[return] F12: Returns to the previous menu.
2nd page
[Max Hold] F7: Compares new and old waveform data items each time a measurement is per-
formed, and displays the larger data item. (Maximum value holding)
[Min Hold] F8: Compares new and old waveform data items each time a measurement is per-
formed, and displays the smaller data item. (Minimum value holding)
[Cumulative] F9: Sets waveform dot data accumulation display mode.
[Over Write] F10: Sets the waveform overwriting mode.
[return] F12: Returns to the previous menu.
[Calibration] F10: Displays the level calibration menu.
[Power Meter Calibration] F7: Performs the level calibration using an internal power meter.
Calibration progress window is displayed during calibration.
[Int. Osc. Calibration] F8: Performs the level calibration using an internal oscillator.
Calibration progress window is displayed during calibration.
[Calibration Cancel] F9: Deletes level calibration data.
[return] F12: Returns to the previous menu.
[Adjust Range] F11: Sets the measurement level ranges (RF power meter range and reference level) to the
status appropriate for measurement signals.
(See paragraph 4.3.1 (2))
[Back Screen] (F12): Displays the previous screen.