4-85
4.3 CDMA Transmitter and Receiver Test --- TX and RX Tester Mode
2nd page
[Reference Level] F3: Sets the reference level of signal input to the MT8801C.
Initial value: 30 dBm
Setting range:42 dBm to 60 dBm (RF Input: Main)
22 dBm to 60 dBm (RF Input: AUX)
Resolution: 1 dB
Function keys
1st page
[Start/Stop] F7: Starts the Access Probe Power measurement. (Note)
[Calibration] F10: Displays the level calibration execution menu.
[Int. Osc. Calibration] F8: Performs the level calibration using an internal oscillator.
Calibration progress window is displayed during calibration.
[Calibration Cancel] F9: Cancels the level calibration data.
[return] F12: Returns to the previous menu.
[Back Screen] F12: Displays the previous screen.
2nd page
[BS Output Level Cal.] F7: Performs the level calibration of the BS Output Level.
[Back Screen] F12: Displays the previous screen.
Note:
Single key or Continuous key can start this measurement, where the Continu-
ous key forcibly makes the single operation.
The single measurement measures the number of frames set in Measuring Pe-
riod. (See paragraph 4.3.6 (1).)