Section 4 Operation
4-90
• Function keys
Main function keys:
[Channel] F1: Changes the channel number.
[Reference Level] F3: Changes the reference level.
[OccBW Ratio] F4: Sets the percentage of total power that resides in the occupied bandwidth.
Initial Value: 99.0 %
Range: 80.0 to 99.9 %
Resolution: 0.1
[RBW] F5: Displays the resolution bandwidth (RBW) menu.
Range: 10 kHz, 30 kHz (Initial value)
Function keys:
[Measure Method] F7: Displays the measurement method (99 % method) selection menu.
[Spectrum] F7: Measures the occupied frequency bandwidth by using the spectrum analyzer
method.
VBW: 30 kHz
SWP Time: When Data Rate is Full : 100 ms
When Data Rate is 1/2 or less : 10 s
[FFT] F8: Measures the occupied frequency bandwidth at high speed by using the fast
Fourier transform (FFT) method.
(Span: 1.95 MHz)
[return] F12: Returns to the previous menu.
[Storage Mode] F9: Displays the storage mode setup menu.
[Normal] F7: Sets normal mode (initial value).
[Average] F8: Sets averaging mode.
The Measure mode is set to Single.
[Average Count] F9: Sets the averaging count.
2 to 9999, Resolution: 1, Initial value: 10
[return] F12: Returns to the previous menu.
[Calibration] F10: Displays the level calibration menu.
[Manual Calibration] F7: Performs level calibration.
During calibration, the window indicating that calibration in progress is dis-
played on the screen.
At Main input: Calibrates using Power Meter and calibration oscillator.
[Calibration Cancel] F8: Deletes the level calibration data.
[return] F12: Returns to the previous menu.
[Adjust Range] F11: Sets the measurement level range (RF power meter range and reference level) to the
status appropriate for measurement signals.
[Back Screen] F12: Displays the previous screen.