4-93
4.3 CDMA Transmitter and Receiver Test --- TX and RX Tester Mode
Function key:
1st page
[Measure Method] F7: Displays the selection menu for the measurement method.
[Spectrum] F7: Uses the spectrum analyzer method.
[Return] F12: Returns to the previous menu.
[Unit] F8: Displays the menu for the power measurement value units.
[dBm] F7: Sets power measurement unit to dBm. (Initial value)
[mW] F8: Sets power measurement unit to mW.
[uW] F9: Sets power measurement unit to uW.
[nW] F10: Sets power measurement unit to nW.
[dB] F11: Sets power measurement unit to dB.
[Return] F12: Returns to the previous menu.
[Storage Mode] F9: Set-up Menu for Storage Mode is displayed.
[Normal] F7: Sets Normal Mode. (Initial value)
[Average] F8: Sets Averaged Mode.
Measure Mode becomes Single.
[Average Count] F9: Sets Averaged Count.
2 to 9999, resolution: 1, initial value: 10
[return] F12: Returns to the previous menu
[Calibration] F10: Displays menu for executing the level calibration.
[Manual Calibration] F7: Executes level calibration.
During calibration, a window indicating the execution of calibration is
displayed.
For Main input: Calibrates by the Power Meter and calibration oscillator.
[Calibration Cancel] F8: Stops the calibration.
[Return] F12: Returns to the previous menu.
[Adjust Range] F11: Sets measurement level ranges (range and reference level of the RF power meter) to
optimal ranges.
[Back Screen] F12: Displays the previous screen.
2nd page
[Recall Template] F7: Displays menu for calling the template to measure spurious emission. (See paragraph
4.3.11)
[Setup Template] F8: Displays setting screen of the template for spurious emission measurement. (See
paragraph 4.3.6 (10) (b)).
[Back Screen] F12: Displays the previous menu.