4-101
4.3 CDMA Transmitter and Receiver Test --- TX and RX Tester Mode
Notes:
1. Forced to single measurement.
2. For a frequency of 1.6 GHz or more, a high pass filter is used for the mea-
surement.
• Function key
Main function key:
[Channel] F1: Changes the Channel number.
[Level] F3: Changes the reference level.
Function key:
1st page
[Spurious Mode] F7: Displays a selection menu of measurement method.
[Spot] F7: Measures levels based on the frequency table. (Initial value)
[Search] F8: Sweeps ±500 kHz centering the spurious frequency on the frequency table,
then finds and displays the maximum level measured.
[Return] F12: Returns to the previous menu.
[Unit] F8: Displays the menu for the power measurement value units.
[dBm] F7: Sets power measurement unit to dBm. (Initial value)
[mW] F8: Sets power measurement unit to mW.
[uW] F9: Sets power measurement unit to uW.
[nW] F10: Sets power measurement unit to nW.
[dB] F11: Sets power measurement unit to dB.
[Return] F12: Returns to the previous menu.
[Calibration] F10: Displays menu for executing the level calibration.
[Manual Calibration] F7: Executes level calibration.
During calibration, a window indicating the execution of calibration is dis-
played.
At Main input: Calibrates by the Power Meter and calibration oscillator.
[Calibration Cancel] F8: Stops the calibration.
[Return] F12: Returns to the previous menu.
[Adjust Range] F11: Specifies measurement level ranges (range and reference level of the RF power meter)
to optimal ranges.
[Back Screen] F12: Displays the previous screen.
2nd page
[Recall Frequency Table] F7: Displays menu for calling the frequency table to measure spurious emission. (See
paragraph 4.3.11)
[Setup Frequency Table] F8: Displays setting screen of the frequency table for spurious emission measurement.
(See paragraph 4.3.6 (11) (b))
[Back Screen] F12: Displays the previous menu.