EasyManuals Logo

Applent Instruments AT2818 User Manual

Applent Instruments AT2818
91 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #14 background imageLoading...
Page #14 background image
14
AT2818/2816A/2816B/2817A/2817/810A User's Guide
Cs
Rs
D=2FCsRs=1/Q
Cp=Cs/(1+D
2
)
Rp=Rs(1+D
2
)/D
2
Q=Xs/RsD=Rs/XsXs=1/2πFCs=2πFLs
2.2.3 Ranging
Auto, Hold and Nominal range. Total 9 Ranges.
About Nominal ranges: (Applent new definition) The AT281x will automatically
select the best range according to the nominal value.
2.2.4 MeasurementSpeed
Fast: 30 readers per second;
Medium 2: 10 readers per second;
Medium 1: 5 readers per second;
Slow: 1.5 readers per second.
Simultaneity, you can specify the averaging factor within the range of 1 to 256 in steps
of 1Averaging Factor to improve performance.
2.2.5 TriggerMode
Include Internal, Manual, External and Bus Trigger.
2.2.6 BasicAccuracy
AT2818 0.05%
AT2816A 0.05%
AT2816B 0.1%
AT2817A 0.05%
AT2817 0.05%
2.2.7 MeasurementRange
Table2‐2 AT2818AT2816AAT2816BAT2817AMeasurementRange
Parameter Measurement Range
L
0.00001H 9999.99H
C
0.00001pF 9999.99mF
RXZ 0.00001Ω 99.9999M
BG 0.01nS 999.999S
D
0.00001 9.99999
Q
0.00001 99999.9
θd -179.999°~179.999°
θr -3.14159 3.14159
Table2‐3 AT2817MeasurementRange
Parameter Measurement Range
L
0.0001nH 9999.9H
C
0.0001pF 9999.9mF
RXZ 0.0001Ω 99.999M
BG 0.01nS 999.99S
D
0.0001 9.9999

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applent Instruments AT2818 and is the answer not in the manual?

Applent Instruments AT2818 Specifications

General IconGeneral
BrandApplent Instruments
ModelAT2818
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals