AT2818/2816A/2816B/2817A/2817/810A User's Guide
Cs
Rs
D=2FCsRs=1/Q
Cp=Cs/(1+D
2
)
Rp=Rs(1+D
2
)/D
2
Q=Xs/RsD=Rs/XsXs=1/2πFCs=2πFLs
2.2.3 Ranging
Auto, Hold and Nominal range. Total 9 Ranges.
About Nominal ranges: (Applent new definition) The AT281x will automatically
select the best range according to the nominal value.
2.2.4 MeasurementSpeed
Fast: 30 readers per second;
Medium 2: 10 readers per second;
Medium 1: 5 readers per second;
Slow: 1.5 readers per second.
Simultaneity, you can specify the averaging factor within the range of 1 to 256 in steps
of 1Averaging Factor to improve performance.
2.2.5 TriggerMode
Include Internal, Manual, External and Bus Trigger.
2.2.6 BasicAccuracy
AT2818 0.05%
AT2816A 0.05%
AT2816B 0.1%
AT2817A 0.05%
AT2817 0.05%
2.2.7 MeasurementRange
Table2‐2 AT2818AT2816AAT2816BAT2817AMeasurementRange
Parameter Measurement Range
L
0.00001H 9999.99H
C
0.00001pF 9999.99mF
RXZ 0.00001Ω 99.9999M
BG 0.01nS 999.999S
D
0.00001 9.99999
Q
0.00001 99999.9
θd -179.999°~179.999°
θr -3.14159 3.14159
Table2‐3 AT2817MeasurementRange
Parameter Measurement Range
L
0.0001nH 9999.9H
C
0.0001pF 9999.9mF
RXZ 0.0001Ω 99.999M
BG 0.01nS 999.99S
D
0.0001 9.9999