85Command Reference
BUS BUS Trigger Mode
Example
SEND> TRIG:SOUR BUS<NL>
Query Syntax
TRIGger:SOURce?
Query Response
{INT,MAN,EXT,BUS}
Example
SEND> TRIG:SOUR? 1
<NL>
RET> INT<NL>
11.16 BIASSubsystem
The BIAS subsystem command group sets the DC BIAS switch to ON or OFF, and sets
the DC bias voltage value.
Figure11‐12 BIASSubsystemCommandTree
Command Syntax
BIAS {OFF,2V,1.5V,EXT}
Example
SEND> BIAS OFF<NL>
Query Syntax
BIAS?
Query Response
{OFF,2V,1.5V, EXTENALINPUT}
Example
SEND> BIAS?
<NL>
RET> OFF<NL>
11.17 FILESubsystem
The FILE subsystem command group executes the file operation.
Figure11‐13 FILESubsystemCommandTree
FILE
:SAVE
:L OAD
<File No.>
<File No.>
:DELete
<File No.>
11.17.1 FILE?
The FILE? query returns the file number used by system.
Query Syntax
FILE?
Query Response
<NR1(0 TO 9):File number>
Example
SEND> FILE?<NL>
RET> 0<NL>
11.17.2 FILE:SAVE
The FILE:SAVE command saves all user settings into current used file.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applent Instruments AT2818 and is the answer not in the manual?

Applent Instruments AT2818 Specifications

General IconGeneral
BrandApplent Instruments
ModelAT2818
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals