EasyManua.ls Logo

Applent Instruments AT2818 - Disp:line; Function Subsystem; Function[:Impedance][:Type]

Applent Instruments AT2818
91 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
71Command Reference
Example
SEND> DISP:PAGE SYST<NL> //Set to the SYSEMT CONFIG
Query Syntax
DISP:PAGE?
Query Response
<page name><NL>
Example
SEND> DISP:PAGE?<NL>
RET> SYST<NL>
11.6.2 DISP:LINE
The :LINE command enters an arbitrary comment line of up to 30 ASCII characters in
the comment field.
Command Syntax
DISP:LINE “<string>”
Parameter
Where, <string> is ASCII character string (30 ASCII characters)
Example
SEND> DISP:LINE “This is a comment.”
11.7 FUNCtionSubsystem
The FUNCtion subsystem command group sets the measurement function, the
measurement range, monitors parameter control.
Figure11‐3 FUNCtionSubsystemTree
11.7.1 FUNCtion[:IMPedance][:TYPE]
The :IMPedance command sets the measurement function.
Command Syntax
FUNC[:IMPedance][:TYPE] <function>
Parameter
Where, <function> is: