83Command Reference
11.14.4 CORRection:SPOT:STA T
The CORRection:SPOT:STATe command sets the LOAD correction function to ON or
OFF.
Command Syntax
CORRection:LOAD:STATe {ON,OFF,1,0}
Parameter
Where, {ON,OFF,1,0} is:
ON, 1 When the function is ON
OFF,0 When the function is OFF
Example
SEND> CORR:LOAD:STATe ON<NL>
Query Syntax
CORRection:LOAD:STATe?
Query Response
{on,off}
Example
SEND> CORR:LOAD:STATe?<NL>
RET> on<NL>
11.14.5 CORRection:SPOT:FREQuency
The CORRection:SPOT:FREQuency command sets the frequency points (FREQ1,
FREQ2 or FREQ3) for the specified frequency point correction.
Command Syntax
CORRection:SPOT:FREQuendy <n>,<value>
Parameter
Where, <n>,<value> is:
n NR1(1,2 or 3):Spot number
value NR1,NR2 or NR3:Frequecny value.
A suffix multiplier can be used with this command. But the unit “Hz” cannot be added.
Example
SEND> CORR:SPOT:FREQ 1,1k<NL>
SEND> CORR:SPOT:FREQ 2,10k<NL>
Query Syntax
CORRection:SPOT:FREQuency? <n>
Parameter
Where, <n> is:
n NR1(1,2 or 3):Spot number
Query Response
<NR3>
Example
SEND> CORR:SPOT:FREQ? 1<NL>
RET> 1.000000e+03<NL>
11.14.6 CORRection:SPOT:OPEN
This command executes the OPEN correction data measure for the specified frequency
(FREQ1, FREQ2 or FREQ3) correction.
Command Syntax
CORRection:SPOT:OPEN <spot number:1 to 3>
Example
SEND> CORR:SPOT:OPEN 1<NL>
11.14.7 CORRection:SPOT:SHORt
This command executes the SHORT correction data measure for the specified frequency
(FREQ1, FREQ2 or FREQ3) correction.
Command Syntax
CORRection:SPOT:SHORt <spot number:1 to 3>
Example
SEND> CORR:SPOT:SHOR 1<NL>
11.14.8 CORRection:SPOT:LOAD
This command executes the LOAD correction data measure for the specified frequency
(FREQ1, FREQ2 or FREQ3) correction.
Command Syntax
CORRection:SPOT:LOAD <spot number:1 to 3>