Page #129 background imageLoading...
Page #129 background image
9.
To prevent damage to the capillary array, inspect each reservoir assembly and
verify that the following conditions are met.
Septa t snugly and ush on the reservoir
Rubber gasket around the edge of the reservoir cap is seated
1
1
Rubber gasket not seated correctly
Plate retainer holes and the septa strip are aligned
1
1
Plate retainer holes and septa holes are not aligned
10.
Dry the reserv
oirs using lint-free wipes.
1
1
Lint-free wipe
11.
Connect the buer reserv
oir plate base cable to the heater outlet inside the
instrument.
IMPORTANT! After installing the buer reservoir, make sure the cable is out of
the way of the autosampler.
1
2
1
Buffer reservoir
2
Buffer position
12.
Place the w
ater and waste reservoirs in the instrument. Load the three reservoirs
in the following order.
a.
Buer reservoir
Chapter 9 Maint
enance
Perform a short-term shutdown
9
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
129

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3730xl and is the answer not in the manual?

Applied Biosystems 3730xl Specifications

General IconGeneral
BrandApplied Biosystems
Model3730xl
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals