SENSseries LB 480
BERTHOLD TECHNOLOGIES GmbH & Co. KG
1 – 77
Volume 1 5 Explosion Protection
1
DQD 507 Rev. 2012-05-22 Page 4
Input Ratings:
100Vac-240Vac, 50/60Hz, 12VA
100Vac-240Vac, 50/60Hz, 12VA
Thermometer Circuit (PT100) Terminals 15 and 16
Type of protection Intrinsic Safety Ex ia IIC Maximum Values:
U
o
= 14.0 V; I
o
= 27.7 mA; P
o
= 97.0 mW; C
i
= 11 nF; L
i
= nil
Maximum permissible external values for common effective reactance’s (C
i
is not considered)
The RTD circuit is electrically connected to the internal supply circuit and the earth.
Open Collector Circuit (Terminal 11, 12)
Type of protection Intrinsic Safety Ex ia IIC Maximum Values:
U
i
= 15.0 V; I
i
= 26.6 mA; P
i
= 100.0 mW; C
i
= 11 nF; L
i
= nil
The open collector circuit is safely electrically isolated from earth and all other circuits.
Hart-current output (Source Mode) (Terminal 17, 18)
Type of protection Intrinsic Safety Ex ia IIC Maximum Values:
U
o
= 25.2 V; I
o
= 101.0 mA; P
o
= 635.0 mW;
Characteristic Linear: C
i
= 3.0 nF; L
i
= 20.0 µH
Maximum permissible external values for common effective reactance’s (C
i
is not considered)
Single reactance’s to table A.2 and figure A.4 or A.6 of 60079-11