114 ATOMLAB 500 DOSE CALIBRATOR
5. Touch the values to change using the <> or <> arrows or key pad to enter the time and
date the dose is to be administrated. Both hours and minutes must be entered if using the
keypad. Touch <Enter> to return to the draw patient dose screen.
NOTE: If the administration time has already passed, you must advance the
administration time to a point in the future. To do this, proceed to the Draw Patient Dose
screen and change the dose administration time.
6. To get a live reading of the dose drawn, select <Verify Activity Drawn>. The updated
Original Kit screen is now displayed.
NOTE: The Administer Dose activity will be displayed in becquerels if this measurement unit
has been previously selected and set up.
NOTE: If you select <Home>, a confirmation screen is displayed noting that data entries will be
lost. Select <Yes> to abort the data without saving. Select <No> to return to the Draw Patient Dose
screen with the patient dose previously entered displayed.
7. Place the drawn dose into the detector. A live reading is now displayed in red showing the
current activity of the patient dose.
NOTE: You can adjust the activity and volume in the syringe, if necessary, and put it back
into the detector to see if you have the correct patient dose. The Administered Dose Activity
for the drawn dose is displayed on the screen. The Administered Dose activity is the decayed
dose activity calculated for what the patient dose will be at the time of administration.
8. If you did not print a label before saving, the screen will now prompt to <Print Label> or
<Skip Printing>. Touch the appropriate choice to proceed.
9. Select <Cancel> to return to the draw patient dose screen without deducting the drawn
dose from inventory. This allows corrections to be made to required dose activity or dose
administration time.
10. Select <Enter> to record the reading and deduct the drawn dose from the original kit.
11. The draw patient dose screen is now displayed. If desired, you can now draw another
patient dose from the kit. Select <Back> to return to the Kit # Data screen (where you can
print a new label for the kit). you can now select <Back> to return to the Kit number
preparation screen.
12. Select <Back> to return to the Nuclear Pharmacy screen.