Biodex Medical Systems, Inc. © 2016 69
Restore Factory Settings:
1. To restore the factory settings to the Setup Isotopes and Kits screen at any time, touch
<Restore Factory Settings>. The system will prompt confirmation.
2. Touch <Yes> to restore the factory settings and return to the Setup Isotopes and Kits screen
with factory settings restored for the isotopes and kit types used for inventory and
Tec-Control. It does not restore factory defaults for other functions within the system.
3. If you have made changes to the inventory isotopes and kits and touch <Enter> the changes
will be recorded and the system returns to the Nuclear Pharmacy screen. If you touch
<Cancel>, the changes are discarded and the system returns to the Dose Calibrator
Configuration Screen.
Low Activity Resolution
Accessed from the Dose Calibrator Configuration screen, this option is used to define the
default display resolution of activity readings. “Auto” is the default setting. To change the
resolution, touch the current setting and then touch your selection from the pop-up list. Touch any
part of the screen while the pop-up list is displayed to cancel without making any changes. The
“Higher” and “Highest” resolution add digits to the activity measurement.
Volume Decimal Digits
This setting is used to determine the number of decimals points, one or two, displayed in
inventory volume readings.
To Toggle Between The One- And Two Decimal Settings:
1. At the Dose Calibrator Configuration screen, touch the Volume Decimal Digits value
displayed and then select either 1 or 2.
2. Touch anywhere on the screen display to enter your selection.