100 ATOMLAB 500 DOSE CALIBRATOR
Enter Tc-99m Into Inventory With Moly
1. From the Nuclear Pharmacy screen, select <Enter Isotope into Inventory>. The Enter Kit #
Data screen is displayed for the currently selected dose calibrator detector. Touch <Select
Detector> if you have multiple detectors and need to change detectors.
NOTE: The kit number is automatically assigned to the lowest number available.
2. Touch <Select Isotope/Kit>. The Select Inventory Isotope And Kit screen is displayed. Touch
Tc-99m to highlight the isotope in the Isotope List. Select the appropriate kit type.
3. Touch <Enter> to return to the Enter Kit # Data screen with the new isotope and kit selected.
4. If a kit has been selected, touch <Enter Lot Number>. Enter the kit lot number with the
on-screen key pad and touch <Enter> to record the lot number and return to the Enter Kit #
Data screen. This could be the generator lot number.
5. Touch <Enter Volume>. Enter the volume with the on-screen key pad. Touch <Enter> to
record the volume and return to the Enter Kit # Data screen.
6. Press <Yes> to perform the Moly Assay. The Moly Assay reading screen is now displayed.
NOTE: If you press <No>, you cannot perform a Moly Assay test.
Figure 5.22. The Enter Kit # Data screen for Tc-99m.