148 ATOMLAB 500 DOSE CALIBRATOR
View Geometry Data
(See Figure 6.18. And 6.19.)
Figure 6.18. The Geometry Test data screen.
The View Geometry Data screen displays the Geometry Tests performed on the dose calibrator.
The data is stored for each detector connected to the system. The display shows the individual
isotope used for a Geometry Test. If multiple isotopes are used, you can view each isotope by
using the <Select Isotope> button. Any Geometry Test outside the acceptable variance is shown
in red. You can view previous Geometry Tests using <Earlier Date> or <Later Date> buttons.
• Touch <Back> to return to the QA Dose Calibrator screen.
• Touch <Select Detector>to select a different detector, if multiple detectors are connected.
• Touch <Select Isotope> to view the Isotope Lists and select the isotope to view for which a
Geometry Test has been previously performed.
• Touch <Print> to print the Geometry Test data displayed.
To Delete Data:
1. Touch <Delete Data> and use the keypad to enter your password. Touch <Enter> to advance
to the Geometry Test data screen.
2. Use <Earlier Date> or <Later Date> to scroll to the test to delete.
3. Touch <Delete Data>. The delete data confirmation screen is displayed. Touch <OK> to
delete and return to the Geometry Test data screen.