Unpacking Instructions
Shipping Cartons
The Atomlab 500 Dose Calibrator and its respective components are carefully packed in two
cartons. One carton contains the display unit and accessories; the other carton contains the
detector unit. Both are shipped in an over-pack.
NOTE: Prior to removing any of the cartons from the shipping box, visually inspect the box and
the inside packing for damage. If any damage is noted, do not unpack the contents. Notify the
carrier immediately so that a claim can be made if necessary, and contact your Biodex Medical
Systems Sales Representative. If there is no visible damage, remove the cartons and place them on
a table or to the side. We suggest removing the display unit carton first.
NOTE: Before deciding on a location to set up your Atomlab for operation, be sure to
read Chapter 2 on installation. This chapter notes operating and environmental
requirements, and offers several tips on choosing a location for your dose calibrator.
The Display Unit Carton:
In addition to the display unit itself, this carton contains the Atomlab's Operation Manual, all the
necessary connection and power wires needed to run the system, and a report of calibration.
The Detector Unit Carton:
CAUTION: The carton labeled DETECTOR UNIT is very heavy, ~38 pounds. Severe damage
to the unit may result if this carton is dropped. Remove the detector from the carton and
carefully place it on the table or other counter where you will set up and use your
Atomlab unit. Make certain the tabletop will support a minimum of 300 pounds. We
suggest that you place the carton on its side so that you can slide the Detector Unit out of the
carton onto its side. Now, stand the detector up onto its base. Inserted into the
Detector Well should be a Detector Well Liner.
ATTENTION: Le carton étiqueté UNITÉ DE DÉTECTION est très lourd, ~38 livres. Cet appareil
Peut subir de sérieux dommages s’ii est échappé. Retirez le détecteur de son carton et utiliserez le
avec prudence sur la table ou le computoir où vous installerez et utiliserez l’unité Atomlab.
Assurezvous que la table supportera un minimum de 300 livres. Nous suggérons que vous
placiez le carton sur son côté afin que vous puissiez faire glisser L’UNITÉ DE DÉTECTION du
carton à
sa base. Dans la chambre du calibrateur vous devriez trouver un manchon pro-
tecteur. Un échantillon en plastique est emballé avec l’unité de présentation.
Save The Packing Material
Do not discard the packing materials, cartons or shipping boxes.
This material was designed especially for the safe shipment of your Atomlab 500 Dose
Calibrator. You will need them to return any of the Atomlab components in the future for
updates, calibration or repair.