108 ATOMLAB 500 DOSE CALIBRATOR
Kit Preparation
(See Figures 5.31 - 5-35.)
Figure 5.31. The Enter New Kit Sample Data screen.
This procedure is normally used to prepare (create) a new kit from an inventory sample of Tc-99m.
To Prepare A New Kit From An Existing Sample:
1. From the Nuclear Pharmacy screen select <Kit Preparation>. The Inventory Kit List screen is
displayed with listings in sample number order. All sample data is current, with the
volumes, Moly readings and activity adjusted when a dose is drawn. The activities are
constantly updated using half-life calculations. If the activity is close to zero for any sample,
the sample line is displayed in red to remind the user to delete it.
2. Press to highlight the isotope/kit from which to draw activity to prepare a new kit. Select
<Enter> to display the Kit # preparation screen.
NOTE: The system automatically assigns a kit number for the new kit preparation and
transfers the Moly concentration information if there was Moly information in the original kit.
3. Touch Select Inventory Kit>. The Select Inventory Kit screen is displayed. Highlight the kit type
desired. If necessary, use <▲> or <▼> to scroll to the appropriate kit type you are making.