Enter Tc-99m Into Inventory Without Moly
1. From the Nuclear Pharmacy screen, select <Enter Isotope into Inventory>. The Enter Kit #
Data screen is displayed for the currently selected dose calibrator detector. Touch <select
detector> if you have multiple detectors and need to change detectors.
NOTE: The kit number is automatically assigned to the lowest number available.
2. Touch <Select Isotope/Kit>. The select inventory and isotope and Kit screen is displayed.
Touch Tc-99m to highlight the isotope in the Isotope List. Select the appropriate kit type.
3. Touch <Enter> to return to the Enter Kit # Data screen with the new isotope and/or kit selected.
4. After selecting the isotope and kit, touch <Enter Lot #>. Enter the kit lot number with the
on-screen key pad and touch <Enter> to record the lot number and return to the Enter Kit #
Data screen. This could be the generator lot number.
5. Touch <Enter Volume>. Enter the volume with the on-screen key pad. Touch <Enter> to
record the volume and return to the Enter Kit # Data screen.
Figure 5.28. The Enter Kit # Data screen for Tc-99m.
6. Press <No> to proceed without performing a Moly Assay.
7. Touch <Zero Background> to begin a background count lasting up to 100 seconds if necessary.
8. When the background time reaches zero, the system returns to the previous screen.
9. The <Record Reading> button should now begin to flash. Place the vial or syringe of Tc-99m
into the detector and touch <Record Reading>.
NOTE: Touch <Print Label> if a label is desired.