EasyManua.ls Logo

biodex ATOMLAB 500 - Sealed Source Setup

biodex ATOMLAB 500
248 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
!
Biodex Medical Systems, Inc. © 2016 59
To Set Acceptance Variance:
1. From the Dose Calibrator Configuration screen, touch <Set Acceptance Variance>.
2. Enter your password and touch <Enter>. The set Acceptance Variance screen is displayed.
3. Touch to highlight the test for which you wish to change the default variance. Then use the
<> or <> keys to increase or decrease the value. Variance is expressed as +/- % for
Constancy, Geometry, Linearity, Decay and Attenuator tests.
4. Touch <Enter> to confirm your changes and return to the dose calibration configuration
screen, or touch <Cancel> to return to the dose calibration configuration screen without
making changes.
NOTE: If you make changes and press <Cancel>, a “Modifications have been made, are you sure you
want to lose the changes?” prompt is displayed. Press <No> to keep the changes, followed by <Enter>
on the Set Acceptance Variance screen. Press <Yes> if you do not want to make the changes and the
system returns to the Dose Calibrator Configuration screen.
Sealed Sources Setup
(See Figures 4.19 - 4.21.)
Figure 4.19. The Sealed Source List. Note that Cs-137 is marked as the default (>>) in this example.

Table of Contents

Other manuals for biodex ATOMLAB 500

Related product manuals