146 ATOMLAB 500 DOSE CALIBRATOR
View Expanded Constancy Data
(See Figure 6.16.And 16.17)
Figure 6.16. The Expanded Constancy Test data screen.
The View Expanded Constancy Data screen displays the expanded constancy tests performed on
the dose calibrator. The data is stored for each detector connected to the system. The
display shows the isotope used for the expanded constancy test, and the test results for all the
isotopes listed for the expanded constancy tests, by the date performed. Any expanded
constancy test outside the acceptable variance is shown in red.
• Touch <Back> to return to the QA Dose Calibrator screen.
• Touch <select DET/SRCE> to bring up the screen to select a different detector or source.
• Touch <Select Detector>to select a different detector, if multiple detectors are connected.
• Touch <Select Sealed Source> if a different sealed source is desired.
• Touch <Delete Data> to delete data.
Use <Earlier Data> or <Later Date> to change the date range displayed.
To Delete Data:
1. Touch <Delete Data> and use the keypad to enter your password. Touch <Enter> to advance
to the Expanded Constancy Test data screen.
2. Touch to highlight the data to delete. Touch <Delete Data>. The Delete Data Confirmation
screen is displayed. Touch <OK> to delete and return to the Expanded Constancy Test data
screen. Touch <Cancel> to return to the Expanded Constancy Test data screen.