EasyManua.ls Logo

biodex ATOMLAB 500 - Page 137

biodex ATOMLAB 500
248 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
!
Biodex Medical Systems, Inc. © 2016 137
Figure 6.8. A Geometry Test with data.
To Perform A Geometry Test:
1. From the QA Dose Calibrator screen touch <Geometry Test>. The Geometry Test screen is
now displayed.
2. If only one detector is connected, skip to step 3. If more than one detector is connected,
touch <Select Detector> and select the detector to test. Zero Background if necessary.
3. The isotope for the Geometry Test always defaults to the first isotope on the button list,
usually Tc-99m. If a different isotope is to be used, touch <Select Isotope> and then select the
desired isotope as previously described.
4. Touch <Set Test Parameters>. The parameters screen is now displayed.
5. Set the Container Description by touching the <Type>, <Size> and <Material>
icons to scroll though the choices. <Type> adjusts syringe or vial, <Size> adjusts from 1 to
30 ml, <Material> is glass or plastic.
6. Touch to highlight volume 1 and then use the <> and <> arrows to adjust the value.
then touch volume 2 and the system automatically increases by the appropriate volume. If
you want a different increase, use the <> and <> arrows to adjust the volume. Touch
volume 3 and the system automatically increases by the appropriate volume. Use the <>
and <> arrows to adjust the value. Continue entering the volumes until you reach the
maximum volume for which you will test.
7. Use the <> and <> arrows to set the normalization volume.
8. Touch <Enter> when all parameters are set to return to the Geometry Test screen.
9. Touch <Zero Background> to zero the background reading, if necessary. The Background
live reading is displayed next to the first volume.

Table of Contents

Other manuals for biodex ATOMLAB 500

Related product manuals