Biodex Medical Systems, Inc. © 2016 153
3. The average decay method described decay corrects each reading, then averages these
calculated activities, and then divides the average by the time corrected reading to
determine a correction factor. If the correction factor is between certain values, the
calibrator is linear.
See Appendix J for additional information.
Linearity Testing Options Include:
• Automated Decay Test
• Semi-Automated Decay Test
• Manual Decay Test
• Lineator Initial Test
• Lineator Test
• Calicheck Initial Test
• Calicheck Test
• Decay Test data
• Lineator Data
• Calicheck Test data
To Access The Linearity Testing menu:
1. From the QA Dose Calibrator screen touch <Linearity Testing>. The Linearity Testing screen
is now displayed.
2. Touch to select the desired function.