Biodex Medical Systems, Inc. © 2016 179
3. Place the Tc-99m vial into the Lineator Tube 1. Place Tube 1 into the detector. The activity
for Tube 1 is now displayed as a live reading.
4. Once the reading has stabilized, touch <Record Reading>. The reading for Tube 1 is recorded
and the line for Tube 1 & 2 is now displayed as a live reading.
5. Touch <Record Reading> to record the reading for Tube 1 & 2.
6. Continue taking readings by touching <Record Reading> after each new tube or set of tubes
is placed in the detector.
7. Touch <End Test> to end the test.
8. Touch <Save Data> to save the test just performed. The Save Lineator Test data screen is
now displayed.
9. If the appropriate staff member is not displayed, touch <Staff Member>. The Staff Member
List is displayed. Touch to highlight any listed person. (If you have previously selected a
staff member, the system will default to that person on the save Lineator Test data
screen.) Then touch <Save Data> again to save the data in memory.
10. Touch <Enter> to select and return to the save Lineator Test data screen.
11. Touch <Save Data> to save the data.
• Touch <Back> twice to return to the Linearity Testing screen.
• Touch <count>to advance to the <Measure>.
• Touch <View Data> to view the linearity tests stored in the system.
NOTE: The Lineator Test must be performed in less than six minutes or the test must be
redone because of decay. See the Lineator instructions.
Figure 6.49. The Save Lineator Test data screen.