!
Biodex Medical Systems, Inc. © 2016 183
6. Continue taking readings by touching <Record Reading> after each new tube is placed over
the Black Tube in the detector (be sure to remove the previous colored Tube before adding
each new tube to the Black Tube).
7. Touch <End Test> to end the test.
8. Touch <Save Data> to save the test just performed. The save Calicheck Initial Test data
screen is now displayed.
9. If the appropriate staff member is not displayed, touch <Staff Member>. The Staff Member
List is displayed. Touch to highlight any listed person. (If you have previously selected a
staff member, the system will default to that person on the save Calicheck Initial Test
data screen.) Then touch <Save Data> again to save the data in memory.
10. Touch <Enter> to select and return to the save Calicheck Initial Test data screen.
11. Touch <Save Data> to save the data.
• Touch <Back> twice to return to the Linearity Testing screen.
• Touch <Measure> to advance to the Measure screen.
• Touch <View Data> to view the Calicheck linearity tests stored in the system.
• Touch <Print> to print the initial Calicheck Test data.