186 ATOMLAB 500 DOSE CALIBRATOR
5. Touch <Record Reading> to record the reading for the Black and Red Tube.
6. Continue taking readings by touching <Record Reading> after each new tube is placed in
the detector (be sure to remove the previous colored Tube before adding each new tube to
the Black Tube).
NOTE: Make sure you have the correct tube combination, as highlighted on the display.
7. Touch <End Test> to end the test.
NOTE: You can only record the same tube combinations as recorded for the Calicheck Initial
Test.
8. Touch <Save Data> to save the test just performed. The Save Calicheck Test data screen is
now displayed.
9. If the appropriate staff member is not displayed, touch <Staff Member>. The Staff Member
List is displayed. Touch to highlight any listed person. (If you have previously selected a
staff member, the system will default to that person on the save Calicheck Initial Test
data screen.) Then touch <Save Data> again to save the data in memory.
10. Touch <Enter> to select and return to the save Calicheck Test data screen.
11. Touch <Save Data> to save the data.
• Touch <Back> twice to return to the Linearity Testing screen.
• Touch <Measure> to advance to the Measure screen.
• Touch <View Data> to view the linearity tests stored in the system.
• Touch <Print> to print the Calicheck linearity test report shown on the screen.