Biodex Medical Systems, Inc. © 2016 51
The Isotope List screen offers six functions related to the maintenance of the list of isotopes.
From the Dose Calibrator Configuration screen, touch <Isotope List> and enter your password
to display the Isotope List screen. At this screen you can Select Current Isotope, Setup Routine
Isotopes, Setup Button Isotopes, Add Isotopes, Remove Isotopes and Change Dial Values.
Select Current Isotope
This option can be used to select an isotope that is not displayed on the Measure screen List 1
or List 2. A full Alphabetical list of Isotopes, the Routine Isotopes List, and Custom Isotopes
with which the Atomlab 500 is designed to work, is provided on the Select Current Isotope
screen.
To Select Any Isotope From The Full Isotope Lists:
1. From the <Home> screen, select <Utilities>. The Utilities screen is now displayed.
2. Select <Dose Calibrator Configuration>. The Dose Calibrator Configuration screen is now
displayed.
3. Select <Isotope List>.
4. Enter your password and touch <Enter>. The Isotope List screen is now displayed.
5. Touch <Select Current Isotope>. The Select Current Isotope screen is displayed.
6. Use the <▲> or <▼> keys to scroll through the list until the desired isotope is displayed.
7. Touch the desired isotope to highlight your selection.
8. Touch <Enter> to select the desired isotope and return to the Isotope List screen with the
new isotope selected, or touch <Cancel> to return to the Isotope List screen without making a
new selection.