P1 Patient Monitor User’s Manual
8-33
Problem Corrective Actions
information, see 8.4.1 Preparing the Patient Skin.
3. Check electrode application sites. Avoid bone or muscular area.
4. Check that electrodes are not dry or used for a prolonged time. Replace
with fresh and moist electrodes if necessary.
No ECG Waveform
1. Check that the ECG gain is not set too low. Adjust the gain as required.
For more information, see 8.5.5 Setting ECG Waveforms.
2. Check that the leadwires and patient cables are properly connected.
Change cable and leadwires.
3. Check that the patient cable or leadwires are not damaged. Change
them if necessary.
Base Line Wander
1. Check for excessive patient movement or muscle tremor. Secure
leadwires and cable.
2. Check that electrodes are not detached or dry and replace with fresh and
moist electrodes if necessary. For more information, see 8.4.1 Preparing
the Patient Skin.
3. Check for ECG filter setting. Set ECG Filter mode to Monitor.
NOTE
:
Physiological alarm and technical alarm information refer to D Alarm