P1 Patient Monitor User’s Manual
10-12
10.8. Setting PR
10.8.1. Setting PR Alarm
You can set PR alarm by following these steps:
1. Select the SpO
2
parameter area or waveform area to enter the【SpO
2
】menu.
2. Select【PR Alarm】submenu.
3. Set alarms as needed.
10.8.2. Setting QRS Volume
If the alarm source is set to PR, the QRS tone is derived from PR measurements. To
set the QRS volume, follow this procedure:
1. Select the SpO
2
parameter area or waveform area to enter the【SpO
2
】menu.
2. Select【PRSetup】submenu.
3. Set【QRSVolume】to the appropriate value.
If the SpO
2
value is effective, the monitor also adjusts the QRS tone (Pitch tone)
according to the SpO
2
value. For information, see 10.7.8Setting Pitch Tone.
10.8.3. Setting PR Source
Current pulse source is displayed in the PR parameter area. The PR from current pulse
source has the following characteristics:
PR is monitored as system pulse and generates alarms when you select PR as
the active alarm source;
PR is stored in the monitor’s database and reviewed in the graphic/tabular
trends; in trend graphs, as the PR curve is in the same color with that of the
PR source, it is unlikely to distinguish the PR source.
To set which pulse rate as PR source, follow this procedure:
1. Select the SpO
2
parameter area or waveform area to enter the【SpO
2
】menu.
2. Select【PRSetup】submenu.
3. Select【PR
Source】,and select a suitable PR source in the drop-down list.
The drop-down list of 【PR
Source】displays the currently valid PR source from top
to bottom according to the priority level. When you select【Auto】, the system will
automatically select the first option in the list as the PR source. If the PR source you set