P1 Patient Monitor User’s Manual
17-5
The steps of setting the ADT gateway are as below:
1. Select 【 Main Menu 】 quick keyfrom 【 System 】 column to select
【Maintenance】input maintenance passwordEnter.
2. Select 【Network Setup】 submenu 【ADT】 submenu.
3. Set the IP address and port of the ADT gateway.
【ADT Query】 is switched on by default. You can load patient information to
the monitor from the ADT server only when this function is enabled.
17.3. Network Printing
You can print the patient information and data through network printing.
17.3.1. Setting the Print Server
To set a network print server, please to do it follow below steps:
1. Select 【 Main Menu 】 quick keyfrom 【 System 】 column to select
【Maintenance】input maintenance passwordEnter.
2. Select 【Print】 submenu 【Printer】 submenu.
3. Set 【Print Server IP】.
17.3.2. Setting Patient Information
You can customize patient information that appears on the printed reports.
17.3.2.1. Setting Patient Information on ECG Reports
You can set the patient information on ECG reports by the following steps:
1. Select 【 Main Menu 】 quick keyfrom 【 System 】 column to select
【Maintenance】input maintenance passwordEnter.
2. Select 【Print】【ECG Report】.
3. On the right of 【ECG Report】, select the desired patient information items.
Patient ID, Patient name, Age and Gender are displayed on the ECG report by
default.