P1 Patient Monitor User’s Manual
12-5
NOTE
:
If NIBP measurement fails, "XX" is displayed; if NIBP measurement is not
taken, "--" is displayed
Outlined NIBP numerics indicate that the measurement exceeds the set time.
So these NIBP values are not recommended for reference.
12.6. Preparing for NIBP Measurements
12.6.1. Preparing the Patient for NIBP Measurements
In normal use, perform NIBP measurement on a patient who is in the following
position:
Comfortably seated
Legs uncrossed
Feet flat on the floor
Back
and arm supported
NOTE:
It is recommended that the patient calms down and relaxes as much as possible
before performing the measurement and that the patient do not talk during the
measurement.
It is recommended to have the patient sit quietly for five minutes before taking
the measurement.
Other factors that have been shown to result in an overestimation of blood
pressure are labored breathing, full bladder, pain etc.
12.6.2. Placing the NIBP Cuff
To place the NIBP cuff, follow this procedure:
1. Verify that the patient category setting is correct.