P1 Patient Monitor User’s Manual
7-15
7.12.2. Setting Intubation Status Time
The default intubation time is 2 minutes. To change the time, following this
procedure:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmsubmenuOthersubmenu.
3. SetIntubation Duration.
7.12.3. Exiting the Intubation Status
To exit the intubation status, choose either of the following ways:
SelectIntubation Statusquick key.
From the bottom of the RESP or CO
2
menu to select Exit
Intubation Status button.
SelectMain Menuquick keyfromAlarmcolumn to selectExit
Intubation Status.
7.13. Other Bed Alarm
Enter other bed observation interface, and when the monitored bed monitor has an
alarm triggered, the alarm light and alarm sound are prompted according to the highest
level of all alarms of the current monitor and other bed monitor. You can view and
manage other bed
alarm. The alarm delay time from the device to other bed is2s.
7.13.1. Other Bed Alarm Reset
You can reset other bed alarm on the monitor. The steps to enable it to reset the bed
alarm are as follows:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmsubmenuRemote Viewsubmenu.
3. OpenReset Remote Bed’s Alarms.
And then Bed View Settings (bed number) window Reset Remote Alarm