P1 Patient Monitor User’s Manual
60
Appendix E Cybersecurity
This chapter mainly describes the information related to cybersecurity of the
monitor.
E.1 Operating environment
Hardware environment
Monitor software is only applicable to P1 patient monitor hardware
platform.
Screen: 5.5LCD screens with resolution 720*1280.
Peripherals: nurse call module.
Software environment
Main board: P1MB
Operating system: LinuxLinux-3.2.0 kernel + Busybox filesystem.
Database: sqlite-3.16.2
Network environment
Apply to LAN
E.2 Network data interface
The communication interface between the monitor and the CMS is wired or wireless
Ethernet, using the standard TCP/IP protocol family, and the application layer data format
follows the Central Monitoring System Network Communication Protocol during
transmission.
E.3 User access control mechanism
a User identification method: after entering the authorization password, you
have the corresponding user type setting authority.
b User types: medical personnel, hospital equipment maintenance personnel,
factory maintenance personnel.
c User authority:
1 Authority of medical staff: No password. Automatically enter the
monitoring interface after starting up, and can be routinely set as required.
2 Authority of hospital equipment maintenance personnel: Enter the